Version:
ALL
Operating System(s): ALL
Problem:
You are using a USB-Serial Adapter obtained from DSA or Sealevel
Systems and cannot communicate with the Engine
Controller.
Solution:
First, make sure you have
the drivers for your adapter loaded into your PC according to the instructions
that came with the hardware.
Once the drivers are installed
follow the procedure below to adjust the settings of your adapter so it will
function properly with the diagnostic software.
Windows 2000
/ XP:
1. Open the Control Panel(Start Menu -> Settings -> Control Panel).

2. Double-Click on System.
3. Select
the Hardware tab in the System Properties window.

4. Click
on the Device Manager... button.
5. Expand
the Ports (COM & LPT) tree by clicking the plus sign.

6. Double-Click on "USB-Serial
Port" or "Sealevel USB-Serial Port" (COM?).
7. Select the Port
Settings tab in the properties window.

8. Click on the Advanced...
button.

9. Make sure the Following
Settings are correct:
- Receive (Bytes) is
set to 64
- Transmit (Bytes) is
set to 64
- Latency Timer (msec)
is set to 1
- Minimum Read Timeout
is set to 0
- Minimum Write Timeout
is set to 0
- All Checkboxes (serial
enumerator, etc.) are not checked.
- The Comm Port Number
should be the same as what you have selected in the Diagnostic Software
Configuration Screen.
10. Disconnect the USB-Serial
Adapter from the PC, wait 5 seconds and then reconnect it.
Windows 95/98/ME:
1. Open the Control Panel (Start Menu -> Settings -> Control Panel).
2. Double-Click on System.
3. Select the Device Manager tab in the System Properties window.
4. Expand the Ports (COM & LPT) tree by clicking the plus sign.
5. Double-Click on "USB-Serial Port" or "Sealevel USB-Serial Port" (COM?).
6. Select the Port Settings tab in the properties window.
7. Click on the Advanced... button.
8. Make sure the Following Settings are correct:
- Receive (Bytes) is set to 64
- Transmit (Bytes) is set to 64
- Latency Timer (msec)is set to 1
- Minimum Read Timeout is set to 0
- Minimum Write Timeout is set to 0
- All Checkboxes (serial enumerator, etc.) are not checked
- The Comm Port Number should be the same as what you have selected in the Diagnostic Software Configuration Screen.
9. Disconnect the USB-Serial
Adapter from the PC, wait 5 seconds and then reconnect it.
Windows NT:
1. Open the Control Panel (Start Menu -> Settings -> Control Panel).
2. Double-Click on Ports.
3. Select
the communications port you wish to change.
4. Click on the Settings...
button.
5. Click on the Advanced...
button.
6. Make sure the Following
Settings are correct:
- Receive (Bytes) is
set to 64
- Transmit (Bytes) is
set to 64
- Latency Timer (msec)
is set to 1
- Minimum Read Timeout
is set to 0
- Minimum Write Timeout
is set to 0
- All Checkboxes (serial
enumerator, etc.) are not checked
- The Comm Port Number
should be the same as what you have selected in the Diagnostic Software
Configuration Screen.
7. Select the "OK"
button to save the settings
8. Disconnect the USB-Serial
Adapter from the PC, wait 5 seconds and then reconnect it.